HSY-432G Portable Oil Contamination Particle Size Analyzer
Contact Info
- Add:江苏省宿迁市经济技术开发区富民大道88号A09栋, Zip:
- Contact: 周治华
- Tel:18817660351
- Email:879196395@qq.com
Other Products
The HSY-432G Portable Oil Contamination Particle Size Analyzer is an instrument developed in accordance with national and international standards such as GB/T 18854-2002 (ISO11171-1999), specifically designed for detecting contamination levels in oils. It is suitable for on-site and laboratory contamination testing of hydraulic oil, lubricating oil, shale oil, transformer oil (insulating oil), turbine oil (turbine oil), gear oil, engine oil, aviation kerosene, water-based hydraulic oil, phosphate ester oil, and other oils. It can also be installed in various hydraulic transmission systems, oil purifiers, cleaning machines, and test benches to enable online monitoring of oil cleanliness. It is widely applicable in industries such as aerospace, power, petroleum, chemical, transportation, ports, metallurgy, machinery, and automotive manufacturing.
I. Main Features
lTest principle based on the light obstruction (extinction) method specified by international hydraulic standards
lHigh-precision laser sensor with wide testing range, stable performance, low noise, and high resolution
lHigh-precision bidirectional plunger pump sampling method with adjustable sampling speed and high sampling volume accuracy
lPiping made of 316L and PTFE materials, suitable for detecting various organic solvents and oils
lSuitable for laboratory or on-site measurements; built-in pressure reduction device for online high-pressure measurements
lCan be connected to an external pressure chamber to create positive/negative pressure for testing high-viscosity samples and degassing
lCompatible with various sampling containers such as standard sampling bottles and cups, or direct connection to hydraulic systems to meet different industry testing requirements
lBuilt-in multiple calibration curves, compatible with all commonly used domestic and international standards for calibration
lSupports automatic calibration, manual calibration, and semi-count calibration
lBuilt-in over ten commonly used standards including NAS1638, GJB420A, GJB420B, ISO4406, SAE4059F, SAE4059CPC, GB/T14039, ГОСТ17216, QC/T29104, JB/T9737, DL/T432, SAE749D, HH005-2018, providing test results for all built-in standards in a single measurement
lSupports custom standard testing and allows setting up to 64 detection channels based on customer requirements
lConfigurable with 1,000 particle size channels for particle size analysis
lBuilt-in data analysis system that automatically determines sample grades according to standards, with automatic data processing and printing functions
l7-inch color touchscreen operation with Chinese and English input, freely switchable language interface, preset, input, modify, store, and upload functions for convenient and fast operation
lAllows setting multiple user login accounts, adding new users, setting permissions, and deleting old users
lOptional tablet control with Win10 operating system, allowing operators to edit personalized signatures
lEquipped with RS232 interface for connecting to computers or laboratory platforms for data processing, and USB for data storage
lBuilt-in lithium battery suitable for field operations, no external power supply required
lRotating stand design for freely adjusting the instrument's testing angle
II. Technical Specifications
lLight source: Semiconductor laser
lParticle size range: 0.8–600 μm (calibrated according to ACFTD 1–100 μm or ISO MTD 4–70 μm(c))
lSensitivity: 0.1 μm
lCustom detection channels: 8–64 channels optional, arbitrary particle size settings
lSampling volume: 0.2–1000 ml, with 0.1 mL intervals
lSampling accuracy: Better than ±1%
lSampling speed: 5–80 mL/min
lCleaning speed: 5–80 mL/min
lCleaning volume: Configurable between 0–90 mL
lCounting accuracy: Error less than ±5%
lResolution: ≤10%
lRepeatability: RSD < 2%
lMaximum coincidence error: 12,000–40,000 particles/mL
lOffline detection viscosity: ≤100 cSt (optional pressure chamber for higher viscosity)
lPressure range: Low pressure 0–0.6 MPa, high pressure up to 42 MPa (optional pressure reducing valve)
lOnline detection interval: Arbitrarily configurable
lSample temperature: 0°C–80°C
lOperating temperature: -20°C–60°C
lStorage temperature: -30°C–80°C
lScreen size: 7 inches
lPower supply: AC 100–240 V, 50/60 Hz
lBattery capacity: 5200 mAh
lBattery runtime: 6–8 hours
lDimensions: 310 × 305 × 135 mm
lInstrument weight: 7.5 kg
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 颀高仪器 |
| Spec: | HSY-432G |
| Stock: | 100 |
| Manufacturer: | |
| Origin: | China / Jiangsu / Suqianshi |